Kaman DIT-5200L Bedienungsanleitung Seite 12

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 18
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 11
_
_____________________________________________________________________________
Kaman Precision Products PART NO: 860522-001
www.kamansensors.com
Last Revised 11/16/2012
- 12 -
R112 Gain Ch X
R117 Offset Ch X
R
217 Offset Ch Y
R212 Gain Ch Y
Figure 9 Potentiometer Locations
Note:
Offset (zero) potentiometers R117 and R217 are not normally installed and are not
included in this calibration sequence.
1. Position the target at Null
2. Adjust the offset control for the measuring channel being calibrated to 0+/-0.001V (unless
otherwise specified).
3. Position the target at –FS (negative full scale)
4. Adjust the gain control for the measuring channel being calibrated to the full scale voltage +/-
0.005mV (unless otherwise specified)
5. Position the target at Null
6. Adjust the offset control for the measuring channel being calibrated to 0+/-0.001V (unless
otherwise specified). (See above note if there is no offset control)
7. Position the target at –FS (negative full scale)
8. Adjust the gain control for the measuring channel being calibrated to the full scale voltage +/-
0.005mV (unless otherwise specified).
9. Repeat steps 1-8 as necessary
Seitenansicht 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 18

Kommentare zu diesen Handbüchern

Keine Kommentare