Instruction Manual KD-2300 Noncontacting Displacement Measuring System EC DECLARATION OF CONFORMITY This apparatus, when installed and
KD-2300 Instruction Manual Part I: What is the KD-2300? •••• 7 Part I: What is the KD-2300? System Technology Kaman Precision Products’s KD-2300 fam
8 •••• Part I: What is the KD-2300? KD-2300 Instruction Manual Nonconductive materials intervening between the sensor and target have little or no e
KD-2300 Instruction Manual Part II: Getting Started •••• 9 Part II: Getting Started Basic System Your KD-2300 system consists of these basic compon
10 •••• Part II: Getting Started KD-2300 Instruction Manual !Connecting External Power Supply The KD-2300 operates from a ± 15 Vdc power source. I
KD-2300 Instruction Manual Part II: Getting Started •••• 11 15 VSupply+15 VSupply-165432DVM,Recorder,Oscilloscope,orComputer Interface+-+ Output- O
12 •••• Part II: Getting Started KD-2300 Instruction Manual
KD-2300 Instruction Manual Part III: The Equipment •••• 13 Part III: The Equipment Operating Principle* The system uses a principle of impedance var
14 •••• Part III: The Equipment KD-2300 Instruction Manual Accessories Recommended accessories for ease of system calibration and convenience are li
KD-2300 Instruction Manual Part III: The Equipment •••• 15 Sensor Options
16 •••• Part III: The Equipment KD-2300 Instruction Manual Sensor Specification The table below lists the sensors available for use with the KD-230
KD-2300 Instruction Manual Contents •••• i Contents Introduction 1 Welcome...
KD-2300 Instruction Manual Part III: The Equipment •••• 17 Measuring Range The specified linear measuring range for an inductive system is directly
18 •••• Part III: The Equipment KD-2300 Instruction Manual material (other than the target) that is positioned within 3 times the diameter of the se
KD-2300 Instruction Manual Part III: The Equipment •••• 19 Surface Mounting Surface mounting will not affect a sensor’s calibration assuming there
20 •••• Part III: The Equipment KD-2300 Instruction Manual Flush Mounting Flush mounting in a conductive fixture requires a cut-out area around the
KD-2300 Instruction Manual Part III: The Equipment •••• 21 Target Material The essential requirement for a proper target is that it be conductive.
22 •••• Part III: The Equipment KD-2300 Instruction Manual under the face of a sensor. The diagram on the right shows the corresponding analog volt
KD-2300 Instruction Manual Part IV: Calibration •••• 23 Part IV: Calibration Factory Calibration All systems are factory calibrated; even so, we rec
24 •••• Part IV: Calibration KD-2300 Instruction Manual Offset Offset is the minimum space or gap between the sensor and target. You should positi
KD-2300 Instruction Manual Part IV: Calibration •••• 25 Calibration Controls You will need to calibrate each measuring channel in your system using
26 •••• Part IV: Calibration KD-2300 Instruction Manual Gain Control Gain control affects the change in output of a system in volts due to a given
Contents •••• ii KD-2300 Instruction Manual Part IV: Calibration 23 Factory Calibration ...
KD-2300 Instruction Manual Part IV: Calibration •••• 27 VDCoffsetDISPLACEMENTFull Scale CalibrationVDCoffsetDISPLACEMENTBipolar Output Calibration
28 •••• Part IV: Calibration KD-2300 Instruction Manual 1. Install the sensor in the calibration or application fixture, ensuring that when the ta
KD-2300 Instruction Manual Part IV: Calibration •••• 29 VDCoffsetDISPLACEMENTBipolar Output Calibration Use this method when your application is be
30 •••• Part IV: Calibration KD-2300 Instruction Manual 6. Repeat Steps 3 through 5 as many times as necessary until you reach the desired output
KD-2300 Instruction Manual Part IV: Calibration •••• 31 High Accuracy Band Calibration DISPLACEMENToffsetVDC0High Accuracy BandCalibrationCALIBRATED
32 •••• Part IV: Calibration KD-2300 Instruction Manual 1. Install the sensor in the calibration or application fixture, ensuring that when the ta
KD-2300 Instruction Manual Appendix A: Glossary •••• 33 Appendix A: Glossary Glossary A/D Converter A device that converts an analog voltage to a d
34 •••• Appendix A: Glossary KD-2300 Instruction Manual Measurement The difference between the upper and lower frequency response limits of a Ban
KD-2300 Instruction Manual Appendix B: System Modifications •••• 35 Appendix B: System Modifications Maximum Sensitivity (Gain Control) When increas
36 •••• Appendix B: System Modifications KD-2300 Instruction Manual GCut this trace formaximum sensitivity (gain)LHCut this trace for least sensitiv
KD-2300 Instruction Manual Introduction •••• 1 Introduction Welcome Welcome to the Series 2300 by Kaman — a family of precision, noncontact displace
KD-2300 Instruction Manual Appendix B: System Modifications •••• 37 E1 & E2 inthis areaE3, E4, & E5in this areaLower PC Board FOR SLAVE UNI
38 •••• Appendix C: Troubleshooting KD-2300 Instruction Manual Appendix C: Troubleshooting Checklist for Basic Troubleshooting This troubleshooti
KD-2300 Instruction Manual Appendix C: Troubleshooting •••• 39 Problems with sensor mounting and nonrepeatable fixturing can be sources of troubl
40 •••• Appendix C: Troubleshooting KD-2300 Instruction Manual 1.CAUSE: Intermittent cable. This problem is typified by large output changes (ful
KD-2300 Instruction Manual Appendix C: Troubleshooting •••• 41 5.CAUSE: A malfunctioning voltmeter. SOLUTION: Check the system output voltage with
2 •••• Customer Service Information KD-2300 Instruction Manual Customer Service Information Should you have any questions regarding this product, pl
KD-2300 Instruction Manual Service Information •••• 3 Kaman Products Standard Limited Warranty Products of Kaman Precision Products are warranted
4 •••• Service Information KD-2300 Instruction Manual !Cautions and Safeguards This information contains important cautions and safeguards associate
KD-2300 Instruction Manual Service Information •••• 5 !All users, despite their level of expertise, should do the following before attempting to use
6 •••• Service Information KD-2300 Instruction Manual Application Information The system speeds gaging and precision measurements by eliminating er